Automatic chlorofluoride analyzer

Short description:

Fully automatic chlorofluoride tester, coal chlorofluoride tester is mainly used...


   baioukeji@gmail.com

   +86-155-6580-8039

   Whatsapp+86-15565808039

Product Detail

Technical Parameters

Fully automatic chlorofluoride tester, coal chlorofluoride tester is mainly used in coal, electric power, coalification and related teaching, scientific research, quality supervision and other units to detect and analyze fluorine, chlorine and other elements of coal.

Fluoride ion chloride ion analyzer in coal

Functional features of a fully automatic chlorofluoride analyzer:

PID control furnace flow, low temperature drift components, high temperature control accuracy, effectively prolong the service life of silicon carbon tubes.

The evaporation is controlled by solid state relay, and the voltage is continuously adjustable from 0 to 220V.

Equipped with imported digital ion meter, precision automatic titrator, computer program to automatically collect various data and curves. Automatic calculation of measurement results.

It can be titrated automatically, and the titration speed and titration amount can be set by yourself. (optional)

Samples can be automatically sent. (optional)

At the same time, our company can also prepare various titration procedures for users with different usage requirements to achieve computerized titration operation and improve laboratory accuracy.


Parameters of automatic chlorofluoride analyzer:

Temperature control accuracy: 11002 ℃

Heating time: heat up to 11002 ° C within 30 minutes

Power supply: AC220V10%, 50HZ

High temperature hydrolysis time: fluorine 30min; chlorine 30min;

Titration time (excluding hydrolysis time):

Fluorine: 3min (after preliminary sample processing, excluding sample processing time, sample processing time is 30min)

Chlorine: 3min (including sample processing time)

Measurement accuracy: fluorine: 150ug/g15; > 150ug/g10%;

Chlorine: 0.01

Measuring range: Fluorine: 10ug/g-2000ug/g

氯:10ug/g-2000ug/g

Test method: high temperature combustion hydrolysis.

Overall power: ≤ 3.5KW.

The stirring speed can be adjusted continuously, digital display. The computer automatically collects the measurement data and automatically calculates.